Products
Ametek | Jofra

RTCt
Reference Temperature Calibrator
up to 6 preset temperatures for faster calibration across frequent tasks
With a graphical display for stability and DLC monitoring, the RTCt offers real-time insights at a glance, adding visual clarity to the numerical displays of the RTC.
Tailor the interface and name your calibration settings for easy access
Expanded range down to -35°C, ideal for extreme applications.
Up to 5 automated temperature switch runs, versus the single run available in the previous model.
Includes rate control in all modes, ensuring precise temperature adjustments.
RTCt
Reference Temperature Calibrator
RTCt 156 | Temperature range | -32 to 155°C / -26 to 311°F |
---|---|---|
Accuracy | with external STS ref. sensor (model B & C): *± 0.040°C / ± 0.072°F with external STS ref. sensor (model B & C): **± 0.060°C / ± 0.108°F with internal reference sensor: ± 0.10°C / ± 0.18°F |
|
Stability | ± 0.005°C / ± 0.009°F | |
Resolution (user-selectable) | 1° or 0.1° or 0.01° or 0.001° | |
Heating time | -32 to -23°C / -22 to 73°F. . . . . . . . . . . . . 4 minutes
23 to 155°C / 73 to 311°F . . . . . . . . . . . 15 minutes |
|
Cooling time | 155 to 100°C / 311 to 212°F . . . . . . . . . . 4 minutes 100 to 23°C / 212 to 73°F . . . . . . . . . . . . 8 minutes 23 to -24°C / 73 to -11°F . . . . . . . . . . . . 14 minutes -24 to -32°C / -11 to -26°F . . . . . . . . . . . 10 minutes 155 to -32°C / 311 to -26°F . . . . . . . . . . . 36 minutes |
|
Time to stability | 10 minutes | |
Immersion Depth | 160 mm / 6.3 in |
RTCt 157 | Temperature range | -45 to 155°C / -49 to 311°F |
---|---|---|
Accuracy | with external STS ref. sensor (model B & C): *± 0.040°C / ± 0.072°F with external STS ref. sensor (model B & C): **± 0.065°C / ± 0.117°F with internal reference sensor: ± 0.10°C / ± 0.18°F |
|
Stability | ± 0.005°C / ± 0.009°F | |
Resolution (user-selectable) | 1° or 0.1° or 0.01° or 0.001° | |
Heating time | -45 to 23°C / -49 to 73°F. . . . . . . . . . . . . 7 minutes 23 to 155°C / 73 to 311°F . . . . . . . . . . . 17 minutes |
|
Cooling time | 155 to 100°C / 311 to 212°F . . . . . . . . . . 4 minutes 100 to 23°C / 212 to 73°F . . . . . . . . . . . . 8 minutes 23 to -24°C / 73 to -11°F . . . . . . . . . . . . 14 minutes >br> -24 to -32°C / -11 to -26°F . . . . . . . . . . . 10 minutes 155 to -32°C / 311 to -26°F . . . . . . . . . . . 36 minutes |
|
Time to stability | 10 minutes | |
Immersion Depth | 160 mm / 6.3 in |
RTCt 168 | Temperature range | -45 to 165°C / -22 to 329°F (Dry and Liquid) <br> -26 to 160°C / -15 to 320°F (Liquid in Container) |
---|---|---|
Accuracy | with external STS ref. sensor (In Insert): *± 0.045°C / ± 0.081°F with external STS ref. sensor (In Insert): **± 0.070°C / ± 0.126°F with external STS ref. sensor (In Bath): *± 0.045°C / ± 0.081°F with external STS ref. sensor (In Bath): **± 0.060°C / ± 0.108°F with external STS ref. sensor (In Container): **± 0.045°C / ± 0.081°F with external STS ref. sensor (In Container): **± 0.060°C / ± 0.108°F with internal reference sensor (In Insert) : ± 0.18°C / ± 0.32°F with internal reference sensor (In Bath) : ± 0.21°C / ± 0.38°F with internal reference sensor (In Bath) : ± 0.14°C / ± 0.25°F |
|
Stability | ± 0.01°C / ± 0.018°F | |
Resolution (user-selectable) | 1° or 0.1° or 0.01° or 0.001° | |
Heating time | -30 to 23°C / -20 to 73°F. . . . . . . . . . . . . 5 minutes 23 to 100°C / 73 to 212°F . . . . . . . . . . . 10 minutes 100 to 165°C / 212 to 329°F . . . . . . . . . . . 12 minutes |
|
Cooling time | 165 to 100°C / 329 to 212°F . . . . . . . . . . 12 minutes 100 to 23°C / 212 to 73°F . . . . . . . . . . . . 22 minutes 23 to 0°C / 73 to 32°F . . . . . . . . . . . . 13 minutes 0 to -15°C / 32 to 5°F . . . . . . . . . . . 17 minutes -15 to -30°C / 5 to -22°F . . . . . . . . . . . 47 minutes |
|
Time to stability | 30 minutes (Block) <br> 15 minutes (Bath) <br> 30 minutes (Bath in Container) | |
Immersion Depth | 180 mm / 7.1 in (Including insulation plug) 150 mm / 5.9 in (Bath version) |